STΑBLE DYNAMIC .AXXAL FIXATOR
This invention is related to an external fixator in type of removable ball-socket articulated dynamic axial fixator used in treatments of fractured, deformed and shortness of long bones in the branches of Orthopaedics and traumatology of the field of medicine.
External fixators used in medicine are classified in several major groups. Sub-groups where that invention falls, and some examples of works under patent belonging to such groups are as follows:
a-Inarticulated: Orthofix External Fixer
b-Double-articulated ball-socket type: Orthofix External Fixator
c-Double articulated hinged type: Wagner External Fixator, Howmedica monotube External Fixator
In fixators of this type, generally there are clamp parts which would hold both epiphysis separately, two articulation pieces adapted to clamps which provide redressing and adaptation of fracture, body illustrating the length of device and elongation rate thereof, distraction units ensuring extension and shortening.
As distraction units achieve extension and shortening actions in the body as an unit annexed to body, and in the manufacture of this unit certain metal alloys being used with higher resistance and higher density but having less perviousness against X-rays, renders more difficult to fix the articulation by X-ray film. For these reasons, excessive use of X-ray device is required to be able to do articulation fixation,
and thus, both patient and doctor are being exposed to X rays more. Anyhow, the risk of fractured reposition loss is still higher due to insufficient articulation fixation. In addition thereto, super-position event is more frequently encountered in X-ray films.
As annexed unit which has been mounted over body and which provides the modification of body length through fitting its parts one into another, could change the length with the help of the keys and by exercising force, this process which renders difficulties of patient's use, cause of delay fcr adaptation of the patient to device, ana extends the time cf staying in hospital.
Controlled dynamization could not be achieved due to device length could be changed by the rate cf force applied to distraction unit, and the risk of progression in shortness is increased consequently.
Usage cf devices in sizes more than one is necessary due to amount of distraction-ccmpressicn is limited. Weights of annexed devices and additional unit lead to loosening of nails during usage.
Articulation angle in existing fixators is maximum 22° and fixator is not sufficiently functional in fractured repositioning due to insufficiency of articulation angle.
In this connection, purpose of this invention is to provide the fixator which would be used in treatments of fractured, deformed and shortness cf long bones at orthopaedics and traumatology being effective to fix the bone in a straight and
stable manner until the completion of welding, and being capable of doing angularity and controlled dynamization, if needed, further its being functional as much as possible, having enough spaciousness for articulation movement and high modularity, stable, would not cause any reposition loss during application and within time; light, hygienic, pervious for X-rays in sufficient degree, capable fcr easy application, smoothly acceptable by patient and easily usable.
Fixator which is developed for attaining the objectives of this invention is shown in enclosed illustrations which reflect following c-tlooks;
Figure 1- Articulation fixation by using fixator on fractured bond.
Figure 2- Complete section of polyplanar clamped fixator.
Figure 3- Complete section of monoplanar clamped fixator.
Figure 4- A view from a section of polyplanar clamp and articulation globe.
Figure 5- A view from a section monoplanar clamp and of clamp cover.
Figure 6- A view of removable articulation from a section.
Figure 7- A section of distraction unit.
Figure 8- A view of sliding clamp cover.
Parts in figures have been numbered one by one, and the descriptions corresponding to these numbers are given below:
(1) Monoplanar Clamp Cover
(2) Body cf Monoplanar Clamp Cover
(3) Bolt
(4) Pin
(5) Extension bolt
(6) Press spring
(7) Extension nut
(8) Ball
(9) Segment
(10) Tightening cleat
(12) Tightening nut
(13) Articulation globe
(14) Polyplanar clamp body
(15) Eccentric Pin
(16) Polyplanar Clamp Bottom Cover
(17) Polyplanar Clamp Intermediary part
(18) Polyplanar Clamp Upper Cover
(19) Derotation Pin
Figure 1 is the outlook of fixator in the position where polyplanar clamps have been fixed on bone through its special nails (Figure 4).
As is seen in Figures 2 and 3, fixator body (10) is telescopic, and works also by moving within extension belt (5). It is possible to use the body in large, middle and small sizes by selecting extension bolts in different longitudes.
Polyplanar clamp, as is seen in Figures 2 and 4, is composed of two bolts (3), polyplanar clamp body (14), polyplanar clamp bottom cover (16), polyplanar clamp intermediary part (17) and polyplanar clamp upper cover (18), and is allowed to place nails in different plans and various combinations. Fixator with inarticulated polyplanar clamp could be used in bones having a broad clinging surface (like metaphase), angularity of fractured and translation thereof could also be corrected through this clamp.
Monoplanar clamp as seen in Figures 2, 3 and 5, containing Monoplanar Clamp cover (1), Monoplanar clamp body (2) and 2 bolts (3) ensures the placement of nail in single plan, and could be used articulated or inarticulated manner according to requirements .
Figure 6 illustrates removable articulation. When needed, by loosening bolt on the clamp (3) space between clamp and tightening nut is broadened, and by turning eccentric pin (15) tightening cleat (11) is loosened and the globe (15, is become idled. Thus, clamp could be given such angularity around a circle of 360° to a direction up to 30° from the center, so the angle of articulation is being increased and therefore device becomes more functional. In order to fix in desired angle, firstly tightening nut (12) is tightened for taking in the slack of articulation. Thereafter, eccentric pin (15) is tightened with "L-alian" wrench by maximum force. Then, clamp is made to be slided en the body of globe (13) until taking in the slack of articulation, and thus clamp body is leaned against tightening nut. In this way, load is being transferred directly to the body
and articulation could be removed or insufficiency position of articulation in angular fixation could be eliminated. Through taking in the slack of articulation by tightening nut and reducing of height of eccentric pin (15) down to maximum 0.5 mm load arm is shortened, more tightening of globe is ensured, and stability is increased by sliding of eccentric in the course of time with a counter-screw applied to eccentric. Distraction- compression share has also been increased by removal of articulation. 3y the grace of removalableness of articulation, problem of reduction loss due to articulation insufficiency is also eliminated and furthermore, stability is increased. Removableness of articulation causes also to ease production and cost decrease as well. Thus, "removable ball-socket type articulated external fixators" has been added to external fixators as a new sub-group. This method could be used in both clamps.
Figure 2 is an outlook of distraction unit. Distraction unit consisting of extension bolt (5), pressing spring '6), extension nut (7), ball (8), segment (9) and derotation pin (19) is placed within body. Therefore, annexed unit which was using for the same purpose has been removed, and in this way, perviousness cf device for X-rays is increased and radioscopy process becomes easier. At the same time, the risk of loosening in nails is also reduced along with decrease of super-position risk and the weight of device. If extension nut (7) is turned clockwise one tour, then the system is being extended 1.2 mm. Meanwhile, spring ball is seated into its three separate sockets with equal intervals placed onto extension nut with a "click"
sound, an warns the user, as well as ensures rotation stability. Here, each "click" sound corresponds 0.4 mm. For compression, these steps would be repeated vice versa. So, by presenting easiness in use, adaptation of patient becomes easier.
Figure 8 is an outlook of sliding clamp cover. Monoplanar clamp cover (1) is slided towards outside by loosening external bolt (3) fcr smooth application of nail to the bone. After fractured reduction is achieved, external screw is tightened. After internal nail is placed into the bone, cover (1) is locked by being slided. This practice provides easiness in technical usage. In Figure 1, the locked position of sliding clamp cover and the area of movement of cover with the help cf bolt hole design could be seen.
Invention could reliably be used in long bones of extremities. As device has been designed in a modularly structure and could be produced in different sizes, suitable fixator is selected as to the kind, location and length of bone. If articulated fixator is preferred, articulation in flat bones is hold firstly straight and closed, distraction-compression share is adjusted. Long bone is rather taken to traction. Nails at both farthest ends of fixator will be placed intc bone at 90° under sterile conditions. Thus, fractured re-position in front- rear plan is being ensured. In second step, nail is placed at the same plan with first nail and parallel thereto into the hole nearest fractured or farthest hole from previous nail. If while placing the nail, central line of bone is not balanced, fractured bone is pulled downward or upward, and central line is so matched. Same process is applied to opposite side and
fractured ends are being reposed in lateral plan. When necessary, third nails are placed into center part in the same plan and parallel to others. Newness in clamp cover is ensured to ease such application.
After both clamps are applied in the same manner, fractured state is controlled by radioscopy or X-ray. If any angularity is observed, then articulation is opened and correction is made. Distraction-compression rate is adjusted and fixator is locked.